Foreword
The purpose of this test method is to provide a standardized procedure for evaluating the resistance of small electrotechnical products (primarily non-hermetic components) by high and low temperature and humid environmental test chamber.
Scope
This test method applies to accelerated damp heat testing of small electrotechnical products.
Limitations
This method is not suitable to verify external effects for specimens, such as corrosion or deformation.
Test Procedure
1. Pre-Test Inspection
2. Specimen Placement
3.Bias Voltage Application (if applicable)
4. Temperature and Humidity Ramp-Up
5. Test Execution
6. Post-Test Recovery
7. In-Test Measurements (if required)
8. Post-Test Inspection
---
Test Conditions
Unless otherwise specified, test conditions consist of temperature and duration combinations as listed in Table 1.
---
Test Setup
1. Chamber Requirements
2. Chamber Materials
3. Temperature Uniformity
4. Specimen Placement
3. Water Quality
---
Additional Information
Table 2 provides saturated steam temperatures corresponding to dry temperatures (100–123°C).
Schematic diagrams of single-container and double-container test equipment are shown in Figures 1 and 2.
---
Table 1: Test Severity
| Temp. (°C) | RH (%) | Duration (h, -0/+2) |
temperature | relative humidity | Time (hours, -0/+2) | ||
±2℃ | ±5% | Ⅰ | Ⅱ | Ⅲ |
110 | 85 | 96 | 192 | 408 |
120 | 85 | 48 | 96 | 192 |
130 | 85 | 24 | 48 | 96 |
Note: Vapor pressure at 110°C, 120°C, and 130°C shall be 0.12 MPa, 0.17 MPa, and 0.22 MPa, respectively.
---
Table 2: Saturated Steam Temperature vs. Relative Humidity
(Dry temperature range: 100–123°C)
Saturation Temp(℃) |
Relative Humidity(%RH) |
100% | 95% | 90% | 85% | 80% | 75% | 70% | 65% | 60% | 55% | 50% |
Dry Temp (℃) | ||||||||||||
100 | 100.0 | 98.6 | 97.1 | 95.5 | 93.9 | 92.1 | 90.3 | 88.4 | 86.3 | 84.1 | 81.7 | |
101 | 101.0 | 99.6 | 98.1 | 96.5 | 94.8 | 93.1 | 91.2 | 89.3 | 87.2 | 85.0 | 82.6 | |
102 | 102.0 | 100.6 | 99.0 | 97.5 | 95.8 | 94.0 | 92.2 | 90.2 | 88.1 | 85.9 | 83.5 | |
103 | 103.0 | 101.5 | 100.0 | 98.4 | 96.8 | 95.0 | 93.1 | 92.1 | 89.0 | 86.8 | 84.3 | |
104 | 104.0 | 102.5 | 101.0 | 99.4 | 97.7 | 95.9 | 94.1 | 92.1 | 90.0 | 87.7 | 85.2 | |
105 | 105.0 | 103.5 | 102.0 | 100.4 | 98.7 | 96.9 | 95.0 | 93.0 | 90.9 | 88.6 | 86.1 | |
106 | 106.0 | 104.5 | 103.0 | 101.3 | 99.6 | 97.8 | 96.0 | 93.9 | 91.8 | 89.5 | 87.0 | |
107 | 107.0 | 105.5 | 103.9 | 102.3 | 100.6 | 98.8 | 96.9 | 94.9 | 92.7 | 90.4 | 87.9 | |
108 | 108.0 | 106.5 | 104.9 | 103.3 | 101.6 | 99.8 | 97.8 | 95.8 | 93.6 | 91.3 | 88.8 | |
109 | 109.0 | 107.5 | 105.9 | 104.3 | 102.5 | 100.7 | 98.8 | 96.7 | 94.5 | 92.2 | 89.7 | |
110 | 110.0 | 108.5 | 106.9 | 105.2 | 103.5 | 101.7 | 99.7 | 97.7 | 95.5 | 93.1 | 90.6 |
(Additional columns for %RH and saturated temp. would follow as per original table.)
---
Key Terms Clarified:
"Unpressurized saturated vapor": High-humidity environment without external pressure application.
"Steady-state": Constant conditions maintained throughout the test.